Virginia Regulatory Town Hall

Proposed Text

highlight

Action:
Replacement of analytic lab regulation for pharmaceutical ...
Stage: Fast-Track
 
18VAC110-20-810 Laboratory requirements for pharmaceutical processors; testing

A laboratory handling, testing, or analyzing cannabis products for a pharmaceutical processor shall:

1. Ensure independence from all other persons involved in the cannabis industry in Virginia, which shall mean that no person with a direct or indirect interest in the laboratory shall have a direct or indirect financial interest in a pharmacist, pharmaceutical processor, cannabis dispensing facility, certifying practitioner, or any other entity that may benefit from the production, manufacture, dispensing, sale, purchase, or use of cannabis products; and

2. Employ at least one person to oversee and be responsible for the laboratory testing who has earned from a college or university accredited by a national or regional certifying authority at least (i) a master's level degree in chemical or biological sciences and a minimum of two years of post-degree laboratory experience or (ii) a bachelor's degree in chemical or biological sciences and a minimum of four years of post-degree laboratory experience.

3. Obtain a controlled substances registration certificate pursuant to § 54.1-3423 of the Code of Virginia authorizing the testing of cannabis products.

4. Provide proof to the board of accreditation in testing and calibration in accordance with the most current version of the International Standard for Organization and the ISO/IEC 17025 or proof that the laboratory has applied for accreditation in testing and calibration in the most current version of ISO/IEC 17025. Any testing and calibration method utilized to perform a cannabis-related analysis for pharmaceutical processors shall be in accordance with the laboratory's ISO/IEC 17025 accreditation. The accrediting body shall be recognized by International Laboratory Accreditation Cooperation.

a. A laboratory applying for authorization to provide cannabis-related analytical tests for pharmaceutical processors shall receive ISO/IEC 17025 accreditation within two years from the date the laboratory applied for ISO/IEC 17025 accreditation. A laboratory may request, and the board may grant for good cause shown, additional time for the laboratory to receive ISO/IEC 17025 accreditation.

b. A laboratory shall send proof of ISO/IEC 17025 accreditation to the board for cannabis-related analytical test methods for pharmaceutical processors for which it has received ISO/IEC 17025 accreditation no later than five business days after the date in which the accreditation was received.

c. A laboratory may use nonaccredited analytical test methods so long as the laboratory has commenced an application for ISO/IEC 17025 accreditation for analytical test methods for cannabis-related analysis for pharmaceutical processors. No laboratory shall use nonaccredited analytical test methods for cannabis-related analysis for pharmaceutical processors if it has applied for and has not received ISO/IEC 17025 accreditation within two years. The laboratory may request and the board may grant for good cause shown additional time for the laboratory to utilize nonaccredited analytical test methods for cannabis-related analysis.

d. At such time that a laboratory loses its ISO/IEC 17025 accreditation for any cannabis-related analytical test methods for pharmaceutical processors, it shall inform the board within 24 hours. The laboratory shall immediately stop handling, testing, or analyzing Cannabis for pharmaceutical processors.

5. Comply with a transportation protocol for transporting Cannabis or cannabis products to or from itself or to or from pharmaceutical processors.